Test Code TSPU Targeted Stimulant Screen, Random, Urine
Additional Testing Requirements
In most cases, no additional testing is needed after the qualitative targeted stimulant test is performed if the parent drug or metabolites found are consistent with the patient's prescribed medications. However, if an unexpected stimulant is found, confirmatory testing can be requested at an additional charge.
Specimen Required
Supplies: Sarstedt Aliquot Tube 5 mL (T914)
Collection Container/Tube: Plastic urine container
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Collect a random urine specimen.
2. No preservative
Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
LPPS | List prescribed stimulants | No | Yes |
TSTIM | Targeted Stimulant Screen, U | No | Yes |
Method Name
Liquid Chromatography Tandem Mass Spectrometry, High-Resolution Accurate Mass (LC-MS/MS HRAM)
Specimen Type
UrineSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Urine | Refrigerated (preferred) | 14 days |
Frozen | 28 days |
Reject Due To
Gross hemolysis | OK |
Gross icterus | Reject |
Clinical Information
Stimulants are sympathomimetic amines that stimulate the central nervous system activity and, in part, suppress the appetite. Amphetamine and methamphetamine are also prescription drugs used in the treatment of narcolepsy and attention-deficit disorder/attention-deficit hyperactivity disorder (ADHD). Methylphenidate is another stimulant used to treat ADHD. Phentermine is indicated for the management of obesity. All other amphetamines (eg, methylenedioxymethamphetamine: MDMA) are Drug Enforcement Administration scheduled Class I compounds. Due to their stimulant effects, the drugs are commonly sold illicitly and abused. Physiological symptoms associated with very high amounts of ingested amphetamine or methamphetamine include elevated blood pressure, dilated pupils, hyperthermia, convulsions, and acute amphetamine psychosis.
Reference Values
Not detected (Positive results are reported with qualitative "Present" results)
Cutoff concentrations:
Methamphetamine: 100 ng/mL
Amphetamines: 100 ng/mL
3,4-Methylenedioxymethamphetamine (MDMA): 100 ng/mL
3,4-Methylenedioxy-N-ethylamphetamine (MDEA): 100 ng/mL
3,4-Methylenedioxyamphetamine (MDA): 100 ng/mL
Ephedrine: 100 ng/mL
Pseudoephedrine: 100 ng/mL
Phentermine: 100 ng/mL
Phencyclidine (PCP): 20 ng/mL
Methylphenidate: 20 ng/mL
Ritalinic acid: 100 ng/mL
Performing Laboratory

CPT Code Information
G0480
80326 (if appropriate for select payers)
Day(s) Performed
Monday through Sunday
Report Available
3 to 4 daysForms
If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.