Sign in →

Test Code OPTMX Opiate Confirmation, Chain of Custody, Meconium

Additional Tests

Test ID Reporting Name Available Separately Always Performed
COCH Chain of Custody Processing No Yes

Specimen Type

Meconium


Specimen Required


Supplies: Chain of Custody Meconium Kit (T653) includes the specimen containers, seals, and documentation required.

Specimen Volume: 1 g (approximately 1 teaspoon)

Collection Instructions: Collect entire random meconium specimen.

Additional Information:

1. Specimen that arrives with a broken seal does not meet the chain of custody requirements.

2. The laboratory recommends sending chain-of-custody specimens by overnight shipment.


Specimen Minimum Volume

0.3 g (approximately 1/4 teaspoon)

Specimen Stability Information

Specimen Type Temperature Time
Meconium Frozen (preferred) 28 days
  Refrigerated  28 days
  Ambient  14 days

Reject Due To

Grossly bloody Reject; Pink OK

Clinical Information

Opiates are naturally occurring alkaloids that are derived from the opium poppy and demonstrate analgesic effects. Opioids are derived from natural and semisynthetic alkaloids of opium or synthetic compounds(1):

-Codeine is a naturally occurring opioid agonist often incorporated into formulations along with acetaminophen or aspirin to increase its analgesic effect.(2) Codeine is metabolized to morphine and subsequently undergoes glucuronidation and sulfation.

-Morphine is an opioid receptor agonist used for major pain analgesia.(2) It has been shown to distribute widely into many fetal tissues(3) and has been detected in meconium.

-Hydrocodone is a semisynthetic analgesic derived from codeine. Hydrocodone is 6 times more potent than codeine and is prescribed for treatment of moderate-to-moderately severe pain.(2) Hydrocodone undergoes O-demethylation in vivo, forming hydromorphone.

-Hydromorphone, a semisynthetic derivative of morphine, is an opioid analgesic. It is 7 to 10 times more potent than morphine, its addiction liability is similar to morphine.(2)

-Oxycodone, a semisynthetic narcotic derived from thebaine. It is metabolized by O-demethylation, forming oxymorphone.(2)

-Oxymorphone is a semisynthetic opioid derivative of thebaine and is indicated for moderate-to-severe pain.(2)

-Heroin, a semisynthetic derivative of morphine, is rapidly deacetylated in vivo to the active metabolite 6-monoacetlymorphine (6-MAM), which is further hydrolyzed to morphine.(2)

 

Opiates have been shown to readily cross the placenta and distribute widely into many fetal tissues. Opiate use by the mother during pregnancy increases the risk of prematurity and small size for gestational age. Furthermore, heroin-exposed infants exhibit an early onset of withdrawal symptoms compared to methadone-exposed infants. These infants demonstrate a variety of symptoms including irritability, hypertonia, wakefulness, diarrhea, yawning, sneezing, increased hiccups, jitteriness, excessive sucking, and seizures. Long-term intrauterine drug exposure may lead to abnormal neurocognitive and behavioral development as well as an increased risk of sudden infant death syndrome.

 

The disposition of opiates and opioids in meconium, the first fecal material passed by the neonate, is not well understood. The proposed mechanism is that the fetus excretes drug into bile and amniotic fluid. Drug accumulates in meconium either by direct deposition from bile or through swallowing of amniotic fluid. The first evidence of meconium in the fetal intestine appears at approximately the 10th to 12th week of gestation, and it slowly moves into the colon by the 16th week of gestation. Therefore, the presence of drugs in meconium has been proposed to be indicative of in utero drug exposure during the final 4 to 5 months of pregnancy, a longer historical measure than is possible by urinalysis.

 

Chain of custody is a record of the disposition of a specimen to document each individual who collected, handled, and performed the analysis. When a specimen is submitted in this manner, analysis will be performed in such a way that it will withstand regular court scrutiny.

Reference Values

Negative

 

Positives are reported with a quantitative liquid chromatography tandem mass spectrometry (LC-MS/MS) result.

Cutoff concentrations for LC-MS/MS testing:

Codeine: 20 ng/mL

Hydrocodone: 20 ng/mL

Hydromorphone: 20 ng/mL

Morphine: 20 ng/mL

Oxycodone: 20 ng/mL

Oxymorphone: 20 ng/mL

Day(s) Performed

Monday through Sunday

Report Available

2 to 3 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

80361

80365

G0480 (if appropriate)

Method Name

Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

Forms

1. Chain of Custody Request is included in the Chain-of-Custody Meconium Kit (T653).

2. If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.